A rare complication of carotid artery stenting: Displacement of marker ring causing locking of stent and incomplete stent expansion


Creative Commons License

Akgul E., Asungur E. H. , Korur K., Aikirnbaev K., Yaliniz H.

AMERICAN JOURNAL OF NEURORADIOLOGY, cilt.28, ss.1403-1404, 2007 (SCI İndekslerine Giren Dergi) identifier identifier identifier

  • Cilt numarası: 28 Konu: 7
  • Basım Tarihi: 2007
  • Doi Numarası: 10.3174/ajnr.a0562
  • Dergi Adı: AMERICAN JOURNAL OF NEURORADIOLOGY
  • Sayfa Sayısı: ss.1403-1404

Özet

We present a rare case of a complication of placement of a carotid artery stent represented by partial opening of a carotid Wallstent caused by displacement of its metal ring marker, which thus hindered complete expansion of the stent. An intraluminally locked carotid stent necessitated referral of the patient for urgent carotid endarterectomy. A possible reason of this unusual complication can be a manufacturing defect, which, to our knowledge, was not documented previously in open public data bases or on the Internet.